Comparison between 2-D and 3-D ultrasound measurements of nuchal translucency

(Ultrasound Review) Prenatal researchers from St. Joannsspital in Salzburg, Austria, have compared nuchal translucency (NT) measurement using 2-D and 3-D ultrasound.

"The risk for trisomies increases with increasing maternal age and NT," they wrote. Improving measurement accuracy should increase screening sensitivity. The study had three objectives. The first was to assess the feasibility of 3-D NT measurement, secondly to determine whether 3-D was accurate and thirdly to compare 2-D and 3-D results. Their findings were published in the December issue of Ultrasound in Obstetrics and Gynecology.

Comparing 2-D and 3-D ultrasound measurements of nuchal translucency, the researchers found that 3-D ultrasound enabled a correct measurement more often. The difference between the two modes however was not statistically significant.

Two hundred and twenty-nine women with a fetus of crown rump length 45–84 mm were included for study. The mean of three nuchal translucency measurements was calculated using both 2-D and 3-D in the transverse and midsagittal planes. Examination duration was limited to twenty minutes and the mean examination time for 2-D and 3-D was comparable. There was a high correlation between 2-D and 3-D measurements and between midsagittal and transverse measurements. Abnormal cases also underwent 3-D surface reconstruction but this failed to add further diagnostic information.

In 60% of cases, three measurements were possible using 3-D, compared with 51.7% of 2-D cases. A successful measurement was obtained using 3-D in 98.6% of cases compared with 96.8% of cases using 2-D. All transvaginal examinations enabled successful measurement, but the majority (95%) of examinations were performed transabdominally.

Differentiation between the skin line and the amnion proved to be the biggest challenge for transabdominal 2-D examinations. Fetal was the main pitfall in obtaining a measurement using 3-D ultrasound.

"The possibility of rotating a stored volume and inspecting it in three orthogonal planes makes 3-D ultrasound a useful tool for nuchal translucency measurements, especially in doubtful cases," they wrote.

"Comparison between 2-D and 3-D ultrasound measurements of nuchal translucency"
G Clementschitsch et al
Dept of Obstetrics and Gynecology, Prenatal Diagnosis and Therapy Unit, St Johannsspital Salzburg, Austria
Ultrasound Obstet Gynecol 2001(December); 18:475-480